Errors and events during measurement


#error100

the measuring cuvette has not been calibrated”.

The error may occurs when:

  • not proper calibration factors of measurement couvete has set,
  • measurement couvete has not been calibrated yet.

Before repeating measurement:

  • Ensure that the calibration process of the measuring cuvette has been carried out and completed successfully, read article please: calibration of the measurement couvette procedure
  • Check if calibration factors of the measurement couvette are set. In the Statistics tab, under the items Volume A and Volume B, the calibration coefficients are showed. Typical values of item Volume A is in range 3.5 – 4.5ml and item Volume B in range 0 – 1ml

#error110

general water issue

If the problem still occurs, contact Technical Support.


#error111

water level sensor issue

The error may occurs when:

  • photometry circuit is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

If the problem still occurs, contact Technical Support. Share error code.


#error112

can’t take water sample

The error may occurs when:

  • filling pipe isn’t is not placed in the aquarium water
  • the filling pipe is to long (longer than 2m),
  • the Smart Tester is placed too high in relation to the tank from which the water sample is taken,
  • the silicon tubes are clogged or damaged,
  • filling pump is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

  • be sure the silicon tubes are connected right – silicon tubes to the aquarium water connection – IN and the used water connection – OUT on the Smart Tester are connected as for normal use,
  • clean measurement couvette (read this, please cleaning of the measuring couvette),
  • be sure Smart Tester and silicon tubes are connected without any issues and are not coggled,
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of the measuring cuvette )
  • make sure there are no other contaminants in the measuring chamber.

If the problem still occurs, contact Technical Support. Share error code.


#error113

can’t remove water sample

The error may occurs when:

  • draining pump is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

  • clean measurement couvette (read this, please cleaning of the measuring couvette),
  • be sure Smart Tester and silicon tubes are connected without any issues and are not coggled,
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette )
  • make sure there are no other contaminants in the measuring chamber.

If the problem still occurs, contact Technical Support. Share error code.


#error120

drops detection circuit don’t work

The error may occurs when:

  • the drop detection circuit is obscured by incorrect placement of the measuring cuvette,
  • drops detection circuit is damaged.

Before repeating the measurement:

If the problem still occurs, contact Technical Support. Share error code.


#error121

can’t clean dropper of reagent 1

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,
  • chemical reagent dropper or pipes are clogged,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,
  • check if dropper is ok (read this, please installation of the reagent droppers ),
  • clean the dropper manualy,
  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),
  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.


#error122

can’t clean dropper of reagent 2

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,
  • chemical reagent dropper or pipes are clogged,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,
  • check if dropper is ok (read this, please installation of the reagent droppers ),
  • clean the dropper manualy,
  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),
  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.


#error123

can’t clean dropper of reagent 3

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,
  • check if dropper is ok (read this, please installation of the reagent droppers ),
  • chemical reagent dropper or pipes are clogged,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,
  • clean the dropper manualy,
  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),
  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.


#error124

can’t move back reagents to their cartidges

The error may occurs when:

  • drops of reagent fell when pump stop,
  • reagents silicon tubes are reversed in at least one pump.

Before repeating measurement:

  • clean the dropper manualy,

If the problem still occurs, contact Technical Support. Share error code.


#error210

general water issue

If the problem occurs, contact Technical Support.


#error 211

water level sensor issue

The error may occurs when:

  • photometry circuit is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

If the problem still occurs, contact Technical Support. Share error code.


#error212

can’t take water sample

The error may occurs when:

  • filling pipe isn’t is not placed in the aquarium water
  • the filling pipe is to long (longer than 2m),
  • the Smart Tester is placed too high in relation to the tank from which the water sample is taken,
  • the silicon tubes are clogged or damaged,
  • filling pump is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

  • be sure the silicon tubes are connected right – silicon tubes to the aquarium water connection – IN and the used water connection – OUT on the Smart Tester are connected as for normal use,
  • clean measurement couvette (read this, please cleaning of the measuring couvette ),
  • be sure Smart Tester and silicon tubes are connected without any issues and are not coggled,
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette ),
  • make sure there are no other contaminants in the measuring chamber.

If the problem still occurs, contact Technical Support. Share error code.


#error213

can’t remove water sample

The error may occurs when:

  • draining pump is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

  • clean measurement couvette (read this, please cleaning of the measuring couvette ),
  • be sure Smart Tester and silicon tubes are connected without any issues and are not coggled,
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette ),
  • make sure there are no other contaminants in the measuring chamber.

If the problem still occurs, contact Technical Support. Share error code.


#error300

photometry circuit don’t work

The error may occurs when:

  • photometry circuit is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

If the problem still occurs, contact Technical Support. Share error code.


#error321

can’t add reagent 1

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,
  • check if dropper is ok (read this, please installation of the reagent droppers ),
  • chemical reagent dropper or pipes are clogged,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,
  • clean the dropper manualy,
  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),
  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.


#error322

can’t add reagent 2

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,
  • check if dropper is ok (read this, please installation of the reagent droppers ),
  • chemical reagent dropper or pipes are clogged,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,
  • clean the dropper manualy,
  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),
  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.


#error323

can’t add reagent 3

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,
  • check if dropper is ok (read this, please installation of the reagent droppers ),
  • chemical reagent dropper or pipes are clogged,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,
  • clean the dropper manualy,
  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),
  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.


#error410

general water issue

If the problem occurs, contact Technical Support.


#error411

water level sensor issue

The error may occurs when:

  • photometry circuit is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

If the problem still occurs, contact Technical Support. Share error code.


#error412

can’t take water sample

The error may occurs when:

  • filling pipe isn’t is not placed in the aquarium water
  • the filling pipe is to long (longer than 2m),
  • the Smart Tester is placed too high in relation to the tank from which the water sample is taken,
  • the silicon tubes are clogged or damaged,
  • filling pump is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

  • be sure the silicon tubes are connected right – silicon tubes to the aquarium water connection – IN and the used water connection – OUT on the Smart Tester are connected as for normal use,
  • clean measurement couvette (read this, please cleaning of the measuring couvette ),
  • be sure Smart Tester and silicon tubes are connected without any issues and are not coggled,
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette ),
  • make sure there are no other contaminants in the measuring chamber.

If the problem still occurs, contact Technical Support. Share error code.


#error413

can’t remove water sample

The error may occurs when:

  • draining pump is damaged,
  • measuring couvette is dirty,
  • measuring cuvette is incorrectly installed.

Before repeating measurement:

  • clean measurement couvette (read this, please cleaning of the measuring couvette ),
  • be sure Smart Tester and silicon tubes are connected without any issues and are not coggled,
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette ),
  • make sure there are no other contaminants in the measuring chamber.

If the problem still occurs, contact Technical Support. Share error code.


#error424

can’t move back reagents to their cartidges

The error may occurs when:

  • drops of reagent fell when pump stop,
  • reagents silicon tubes are reversed in at least one pump.

Before repeating measurement:

  • clean the dropper manualy,

If the problem still occurs, contact Technical Support. Share error code.


Errors & events during calibration of the measurement couvette

Error in step 4/7

Failed to take a water sample. The entire calibration process must be stopped and started again. Before repeating the calibration process, check:

  • if the silicon tubes have been connected to the Smart Tester and placed in a beaker of water,
  • if the silicon tubes is of the required length,
  • if the measuring cuvette is clean (read this, please cleaning of the measuring couvette ),
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette ),
  • make sure there are no other contaminants in the measuring chamber.

If the problem occurs, contact Technical Support.

The error occurs when:

  • the photometric circuit is defective,
  • the measuring cuvette is dirty,
  • the filling tube is not placed in the aquarium water,
  • the filling tube is too long (longer than 2 m),
  • Smart Tester is placed too high in relation to the beaker from which the water sample is taken,
  • the filling tube is clogged or damaged,
  • the filling pump is defective,
  • measuring cuvette is incorrectly installed.

Error in step 7/7

The specified weights in step 3/7 ,5/7 or 7/7 are not correct. The entire calibration process must be stopped and started again. Before restarting the calibration process, check:

  • if you have a balance with a resolution of at least 0.01g,
  • if the balance has an accuracy class of at least +-0.05g,
  • that the silicon tubes have the required length and are unobstructed.

If the problem occurs, contact Technical Support.

The error occurs when:

  • the measured volume of the measuring cuvette is outside the range <1ml – 10ml> i.e. the weight in step 3/7 minus the weight in step 5/7 is not within the range 1g-10g
  • the volume of water that remains in the measuring cuvette is greater than 1ml. i.e. the weight in step 3/7 minus the weight in step 7/7 is greater than 1g

Errors & events during checking of calibration of the measurement couvette (1)

Error in step 2/3

Failed to take a water sample. The entire process must be stopped and started again. Before repeating the calibration process, check:

  • if the silicon tubes have been connected to the Smart Tester and placed in a beaker of water,
  • if the silicon tubes is of the required length,
  • if the measuring cuvette is clean (read this, please cleaning of the measuring couvette ),
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette ),
  • make sure there are no other contaminants in the measuring chamber.

If the problem occurs, contact Technical Support.

The error occurs when:

  • the photometric circuit is defective,
  • the measuring cuvette is dirty,
  • the filling tube is not placed in the aquarium water,
  • the filling tube is too long (longer than 2 m),
  • Smart Tester is placed too high in relation to the beaker from which the water sample is taken,
  • the filling tube is clogged or damaged,
  • the filling pump is defective,
  • measuring cuvette is incorrectly installed.

Errors & events during checking of calibration of the measurement couvette (2)

Error in step 2/7

Failed to take a water sample. The entire process must be stopped and started again. Before repeating the calibration process, check:

  • if the silicon tubes have been connected to the Smart Tester and placed in a beaker of water,
  • if the silicon tubes is of the required length,
  • if the measuring cuvette is clean (read this, please cleaning of the measuring couvette ),
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette ),
  • make sure there are no other contaminants in the measuring chamber.

If the problem occurs, contact Technical Support.

The error occurs when:

  • the photometric circuit is defective,
  • the measuring cuvette is dirty,
  • the filling tube is not placed in the aquarium water,
  • the filling tube is too long (longer than 2 m),
  • Smart Tester is placed too high in relation to the beaker from which the water sample is taken,
  • the filling tube is clogged or damaged,
  • the filling pump is defective,
  • measuring cuvette is incorrectly installed.

Errors & events during reagents set change

Error in step 3/9

Failed to clean reagent dropper A. Before repeating the step, check:

If the problem occurs contact Technical Support.

The error occurs when:

  • the RO water cartridge has not been connected,
  • the droplet detection circuit is defective,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • the reagent pump is defective,
  • obstructions in measuring chamber interfering with the operation of reagent droplet detection circuit, or measuring cuvette has not been correct installed.

Error in step 4/9

Failed to clean the reagent dropper B. Before repeating the step, check:

If the problem occurs contact Technical Support

The error occurs when:

  • the RO water cartridge has not been connected,
  • the droplet detection circuit is defective,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • the reagent pump is defective,
  • obstructions in measuring chamber interfering with the operation of reagent droplet detection circuit, or measuring cuvette has not been correct installed.

Error in step 5/9

Failed to clean the reagent dropper C. Before repeating the step, check:

If the problem persists contact Technical Support

The error occurs when:

  • the RO water cartridge has not been connected,
  • the droplet detection circuit is defective,
  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,
  • the reagent pump is defective,
  • obstructions in measuring chamber interfering with the operation of reagent droplet detection circuit, or measuring cuvette has not been correct installed.

Error in step 7/9

The reagent droppers could not be cleaned. If the problem occurs then contact Technical Support

The error occurs when:

  • the droplet detection circuit is defective,
  • the droppers are defective,
  • chemical reagent pumps are defective.

Error in step 9/9

Failed to rinse/empty the measuring cuvette. Before repeating the step process, check:

  • if the silicon tubes have been connected to the Smart Tester,
  • if the silicon tubes have the required length,
  • if the measuring cuvette is clean (read this, please cleaning of the measuring couvette ),
  • make sure the measuring cuvette is properly seated in the capsule and installed in the Smart Tester (read this, please installation of measuring cuvette ),
  • make sure there are no other contaminants in the measuring chamber.

If the problem occurs, contact Technical Support.

The error occurs when:

  • the photometric circuit is defective,
  • the measuring cuvette is dirty,
  • the filling tube is not placed in the aquarium water,
  • the filling tube is too long (longer than 2 m),
  • Smart Tester is placed too high in relation to the beaker from which the water sample is taken,
  • the filling tube is clogged or damaged,
  • the filling pump is defective,
  • measuring cuvette is incorrectly installed.